0 to 50 deg C (32 to 122 deg F) RH max 85% non-condensing
Data Logging
6000 pH/ deg C or ORP samples
Analog Output Accuracy
+- 2% Fs
Analog Output Resolution
0.1% Fs
Max. Oscillation Frequency
8 MHz
Power Supply
230 VAC +-10% ; 50/60 Hz
Temperature Compensation
Automatic or Manual, -30 to 130 deg C
HI 504 ORP Digital Controller is a PID, PI, proportional or on/off pH/ORP controller with one or two set points. The measurement configuration settings and control of pH and ORP are saved separately and permits users to switch between pH and ORP without losing settings. The pH channel can be calibrated in 2 calibration points. The instrument has a full auto diagnostic procedure. Sensor Check is also available for pH and ORP probes. The temperature is continuously monitored using a temperature sensor (Pt100 or Pt1000 type) with automatic temperature compensation of pH. One or two analog controller outputs (0-20 or 4-20 mA) can be configured for pH/ORP recording or controlling (only for models with PID), and relays can be used to control the process or be connected with alarm status. Controller status is visable with LED's on the front panel and on LCD. The controllers logging feature can save up to 6000 samples pH/°C or ORP and last 100 error, configuration, calibration and cleaning events. This information can accessible from a PC through RS485 and software. The powerful software has graphing capabilities and can print graphs directly or can be saved as a bitmap. Data can be exported in common spreadsheet formats.